DART Operation Requirements and Recommendations

Power Requirements:

110 VAC @ 60 Hz and 5 Amps. (50 Hz acceptable: must be set in software.)
Europe: Must use isolation transformer to step-down from 230 to 110 VAC. We recommend a Single Phase Type 1 Isolation Transformer from Radio Spares (Part no 260-4272 at www.rswww.com). A specific mains plug will be needed, Part no 235-9950.

Computer:

A computer with Windows XP and one available USB port.

Gas:
Regulator:

Our recommendation is a single stage regulator for each gas tank with the following specs:

Filter:

We recommend the use of a particulate filter on both N2 and He gas lines entering the DART Controller. This helps to ensure no foreign objects (particularly with in-house gas lines) contaminate or block gas flow through the DART. The filter should be placed after the regulator and before the DART Controller.
(Part no: SGD7510-10-P4MM from Specialty Gas Equipment Inc. www.sgd.com)

Flow Meter:

If customers wish to measure the flow rate of DART gas with an external meter, IonSense recommends the following: