  Dynamic Range in Thermal Profiling of Tablets

In order to investigate the dynamic range of to DART for Tablet analysis we used the Temperature Profiling Method to analyze both the Daytime and Nighttime Caplets. Using the 350C profile as the optimum temperature for simultaneous generation of ions, we were able to detect the Chlorpheniramine Maleate present at only 0.5% of the tablet mass.

Tylenol
Night Caplet

  • Acetaminophen
  • -325 mg/caplet

  • Dextromethorphan HBR
  • -10 mg/caplet

  • Phenylephrine HCl
  • -5 mg/caplet

  • Chlorpheniramine Maleate
  • Nighttime Caplet Chart

    In the case of a low resolution mass analyzer we are able to determine the presence of the Chloropheniramine by using the isotope ratio between the Cl 35 and Cl 37. A high resolution accurate mass analysis could serve to confirm the ID.

    Treat Nighttime Caplet Chart

    An alternative method is to change the speed of sampling stage. The faster speed provides further utility for analysis.